Geno™ CL - Clinical Laboratory CLRW Water

The Avidity Science® Geno™ CL series CLRW system is a one-box solution that can deliver up to 200 L/hr. of deionized water with a versatile and intuitive user system that exceeds expectations for a clinical scientist and a high throughput analyzer.

  • A range of flow rates in a ‘one box’ solution with a small footprint to match the requirement of any analyzer system configuration.
  • Unique two-stream RO/EDI cassette technology provides contingency, upgrade capability and minimizes downtime
  • Simple operator interfaces through an Intuitive icon driven control via a high resolution 10” touch screen. 
  • CLSI-CLRW compliant water delivered by whisper-quiet pumps
  • Data logging via a secure link to protected cloud storage
  • Onboard maintenance tutorial videos
  • Passcode protected access and prioritized alarm warnings
  • Low running costs
  • Tool-free access and simple consumable change
  • Common pumps reduce parts inventory
  • By-pass options

Download the Avidity Science Geno CL Brochure

Request a Quote

Availability: In stock
SKU
GENOCL

A Flexible, Scalable And Intuitive Water Purification System To Suit Your Clinical Laboratory Needs Today And Tomorrow

The Avidity Science Geno™ CL series has been designed to change the approach to clinical analyser feedwater for diagnostics laboratories.
It minimizes spatial requirements while reducing the risk of water outage; all in one sophisticated cabinet.

The proven water purification technologies used in the production of CLRW grade water have been expertly housed and are complimented by a wide range of features which ensure the Geno™ CL sets a new standard for the supply to clinical analyzers across the industry.

The Geno™ CL platform utilizes a combination of proven water purification technologies, including Reverse Osmosis (RO), Membrane Degassing and Electrodeionization (EDI) Cassettes, combined with in-built storage and distribution to deliver CLRW direct to the clinical analyzers.
All technologies are easily accessible to the operator enabling simple and timely cartridge replacement

Twin stream technology provides the contingency required in the modern laboratory inside one system. A break tank and integral storage reservoir are both held inside this complete package.


 

Flexibility and Scalability Provided by Unique Electrodeionization-Cassette Technology (EDI-CT)

Avidity Science Geno CL EDI-Cassette Technology (EDI-CT)

  • EDI gives consistent quality with predictable running costs
  • Each cassette holds an EDI cell and degasser
  • A separate cassette holds the RO module and pump allowing the end user to increase capacity with growing demand

A Continuous Water Supply Regardless of Demand

Avidity Science Geno CL EDI-Cassette Technology (EDI-CT)

  • EDI-CT allows duplex production from one system, providing built-in contingency
  • Each stream works independently, ensuring no single point of failure
  • Cassettes can be independently replaced with minimal downtime and intervention

Cloud Based Diagnostics and Data Logging

Avidity Science Geno CL Icon Based Touchscreen            

  • Data logging via secure link to protected cloud storage
  • Remote diagnostics via an Ethernet or 4G connection
  • Common fault output for connection to Building Management System

Easy to use Icon-Based Interface

Avidity Science Geno CL Video Tutorials

  • 10” High Resolution Touchscreen
  • Intuitive, pictorial settings, reducing language barriers
  • Password protected interface
  • Alarm warnings with a two stage, time and date stamp
  • Consumable reminder with on screen video user guides

Simple to Change Consumables

Avidity Science Geno CL | Easy to change consumables

  • No tools needed to replace consumables
  • Easy access door - Change consumables in seconds
  • On-site maintenance saving you valuable time and service callouts

Smart Assistance:

       •   Easy to use on-site maintenance saving you valuable time and service callouts
       •   Video tutorials easily accessed from 10" touchscreen
       •   All main consumables located at front for ease of access

Avidity Science Geno™ CL Technical Specifications

SYSTEM REQUIREMENTS

FEED WATER

GENO™ 50 CL & GENO™ 100 CL / CLD 

GENO™ 150  CLD & GENO™ 200 CLD
Pressure (Bar) 1 to 6 1 to 6
Flow (L/hr) 190 380
Hardness as CaCO3 (ppm) <20 <20
Temperature (°C) <25* <25**
Conductivity (µS/cm)*** 10 to 2000 10 to 2000
Free Chlorine (ppm) <0.2* <0.2**
Carbon Dioxide (ppm)*** <30 <30
pH 4 to 10 4 to 10

* Calculated at operation of 12 hours per day, filter change every 6 months

** Calculated at operation of 12 hours per day, filter change every 3 months

*** FCE (Feed Conductivity Equivalent) means conductivity + (ppm CO2 x 2,66) + (ppm SiO2 X 1,94)


PRODUCT WATER QUALITY

PRODUCT WATER SPECS GENO™ 50 CL GENO™ 100 CL / CLD GENO™ 150  CLD GENO™ 200 CLD
Makeup flowrate (L/hr) @ 20°C 50 100 150 200
Inorganics (MΩ-cm at 25°C) >10 >10 >10 >10
Organics (TOC ppb) <30 <30 <30 <30
Bacteria (CFU/mL) <10 <10 <10 <10
Particles (at µm filter) 0.2 0.2 0.2 0.2
Silica (mg/L) <0.05 <0.05 <0.05 <0.05
Delivery flowrate (L/min @ 1 bar) 7.5 7.5 7.5 7.5


UNIT SPECIFICATIONS

ALL MODELS
Dimensions (HxWxD) 1852 x 750 x 600mm
Working Depth 800mm
Net Weight 182kg
Operating Weight 292kg
Maintenance Access 450 x 800mm
Power Supply 110 - 230V ~; 50/60Hz; 1160W (max)


FLOW CHART

Avidity Science Geno CL Flowchart


UTILITIES & SERVICES

REQUIREMENTS

Feed Water Connection 3/4" Male
Drainage (normal) 3.3L/min
Drainage (overflow) 10L/min


CONSUMABLES & ACCESSORIES

PART NUMBER

DESCRIPTION

TYPICAL ANNUAL REQUIREMENT

TC003 UV Lamp 12 Months
TC005 Air Vent Filter 12 Months
TC041 EDI Cell 50 - 100L/h As Required
TC042 RO Membrane 2 - 3 Years
TC043 20" Carbon Pre-filter 3 Months*
TC044 DI Cartridge 3 Months*
TC045 0.2µm Clamp 6 Months
TC046 Degasser Air Filter 6 Months
TC047 Degasser Membrane 2 - 3 Years

* Depending on water consumption