Type II - DI Water
The Cascada II.I is fully integrated to produce 30 L/hr of Type II water directly from tap water. It dispenses up to 2L/min of both Type II pure water and Type I ultrapure water from a flexible dispenser. Real-time water quality and operating conditions are displayed on the dispensing interfrace. A compact dispensing stand (28x30 cm) offers true flexibility to users to maximize bench space utilization and to locate the dispenser at the most convenient point of use.
Routine system control functions are fully availalbe on the dispenser including "Print Report" for Good Laboratory Practices. This allows for maximum bench space utilization.
Water Type | Type II (DI Water) | Type I (Ultrapure Water) |
Resistivity (MΩ-cm @ 25 °C) | < 5 | 18.2 |
TOC (ppb) | < 30 | < 5 *** |
Output Flow Rates | Up to 2 | Up to 2 |
Bacteria (cfu/mL) | <0.1 * | <0.1 * |
Particles (0.2 µm) | <1/mL * | <1/mL * |
Endotoxins (Eµm/mL) | <0.001 ** | <0.001 Eµm/mL ** |
* With 0.2 µm bacteria removal final filter | ** With postive charged Nylon KA1 endotoxin removal final filter | *** With feed TOC less than 50 ppb
System | RO | Type II | Type I |
Cascada II. I - 5 | 5 L/h | > 2 L/min* | Up to 2 L/min |
Cascada II. I - 10 | 10 L/h | > 2 L/min* | Up to 2 L/min |
Cascada II. I - 20 | 20 L/h | > 2 L/min* | Up to 2 L/min |
Cascada II. I - 30 | 30 L/h | > 2 L/min* | Up to 2 L/min |
* From top on reservoir
Dual Water | 1- 15 MΩ-cm Type 2 & 18.2 MΩ-cm Type 1
Integrated Controls on the Cascada II.I system enable Type II water to be availalbe from the reservoir. Water quality is maintained by optmizing recirculation, UV sterilization and vent filtration to minimize recontamination from CO2 volatile organize compounds, bacteria and particles. The system maximizes purification power at each stage so that the pure and ultrapure water outputs are consistencly high in quality. In addition, unique system monitoring and water leakage protection functions offer peace of mind on your laboratory's pure and ultrapure water supply.
Width |
Height |
Depth |
Dry Weight |
366 mm |
575 mm |
492 mm |
23 kg |
Width |
Height |
Depth |
Dry Weight |
280 mm |
845 mm |
300 mm |
6 kg |
Size |
Width |
Height |
Depth |
Dry Weight |
35 L |
390 mm |
600 mm |
384 mm |
5 kg |
70 L |
390 mm |
900 mm |
384 mm |
7 kg |
105 L |
390 mm | 1200 mm | 384 mm | 9 kg |
Width |
Height |
Depth |
Dry Weight |
220 mm |
463 mm |
380 mm |
7 kg |
Feed Water Requirements | |
Conductivity | < 1500 µS/cm @ 25 °C |
Pressure | 0.5 -6 bar |
Temperature | 5 - 40 °C |
Free Chlorine | < 3 ppm |
Silt Density Index | < 12 |
pH | 4 - 10 |
Electrical Requirements | |
Input Voltage | 100 - 240 V (50-60Hz) |
Power |
<200 VA - Unit 75 VA - Pretreatment |
Temperature | 5 - 40 °C |